HomeProductsRefrigerators and freezers
Refrigerators and freezers · POL-EKO

Stable temperature, proven cold chain.

From pharmaceutical refrigerators for vaccines and medicines to laboratory freezers and ultra-low freezers down to −86 °C, POL-EKO covers the full temperature range for samples and temperature-sensitive products. APICAL helps you match the choice to application, capacity and validation requirements.

Cold chain, mapping and qualification as part of validation
POL-EKO laboratory refrigerators
POL-EKOCooling equipment for laboratories and pharmacies

Temperature is not a setting, it is a quality requirement. The right unit, monitoring and documented validation together protect samples, medicines and vaccines.

APICAL × POL-EKOAn overview of device families, capacities and temperature ranges.
0…+10 °C · 74–1397 l

CHS-C pharmaceutical refrigerators

For medicines, vaccines and light-sensitive preparations. Stainless steel interior, 0.1 °C setting resolution. Each capacity is available with a solid door and with a glass door (suffix A).

CHS-C 1CHS-C 1 A

CHS-C 1 / CHS-C 1 A

Chamber volume
74 l
Working capacity
54 l
Temperature
0…+10 °C
Specification
CHS-C 2CHS-C 2 A

CHS-C 2 / CHS-C 2 A

Chamber volume
157 l
Working capacity
129 l
Temperature
0…+10 °C
Specification
CHS-C 3CHS-C 3 A

CHS-C 3 / CHS-C 3 A

Chamber volume
208 l
Working capacity
171 l
Temperature
0…+10 °C
Specification
CHS-C 4CHS-C 4 A

CHS-C 4 / CHS-C 4 A

Chamber volume
259 l
Working capacity
212 l
Temperature
0…+10 °C
Specification
CHS-C 5CHS-C 5 A

CHS-C 5 / CHS-C 5 A

Chamber volume
312 l
Working capacity
255 l
Temperature
0…+10 °C
Specification
CHS-C 6CHS-C 6 A

CHS-C 6 / CHS-C 6 A

Chamber volume
414 l
Working capacity
339 l
Temperature
0…+10 °C
Specification
CHS-C 500CHS-C 500 A

CHS-C 500 / CHS-C 500 A

Chamber volume
506 l
Working capacity
467 l
Temperature
0…+10 °C
Specification
CHS-C 700CHS-C 700 A

CHS-C 700 / CHS-C 700 A

Chamber volume
641 l
Working capacity
598 l
Temperature
0…+10 °C
Specification
CHS-C 1200CHS-C 1200 A

CHS-C 1200 / CHS-C 1200 A

Chamber volume
1397 l
Working capacity
1318 l
Temperature
0…+10 °C
Specification
0…+15 °C · 74–1577 l

CHL laboratory refrigerators

Single-chamber refrigerators with forced air circulation, Smart or Smart PRO controller and solid or glass door options. Available in C (comfort) and P (premium) versions; the Smart controller has a 4.3" display and LAN, while Smart PRO has a 7" display with LAN and Wi-Fi for monitoring through LabDesk software.

CHL 500

CHL 500

Chamber volume
506 l
Working capacity
475 l
Temperature
0…+15 °C
Specification
CHL 700

CHL 700

Chamber volume
641 l
Working capacity
601 l
Temperature
0…+15 °C
Specification
CHL 1200

CHL 1200

Chamber volume
1397 l
Working capacity
1318 l
Temperature
0…+15 °C
Specification
CHL 1450

CHL 1450

Chamber volume
1564 l
Working capacity
1479 l
Temperature
0…+15 °C
Specification
−25 / −40…0 °C · 85–314 l

Laboratory freezers

ZLN 85 and ZLN-T with natural convection, and ZLW-T with forced air circulation for better uniformity and faster cooling of the load.

ZLN-T 125

ZLN-T 125

Chamber volume
138 l
Working capacity
127 l
Temperature
-40…0 °C
Specification
ZLN-T 200

ZLN-T 200

Chamber volume
217 l
Working capacity
202 l
Temperature
-40…0 °C
Specification
ZLN-T 300

ZLN-T 300

Chamber volume
314 l
Working capacity
293 l
Temperature
-40…0 °C
Specification
ZLW-T 200

ZLW-T 200

Chamber volume
217 l
Working capacity
172 l
Temperature
-40…0 °C
Specification
ZLW-T 300

ZLW-T 300

Chamber volume
314 l
Working capacity
263 l
Temperature
-40…0 °C
Specification
−86…−50 °C · 259–482 l

ZLN-UT ultra-low freezers

Vacuum VIP insulation, natural convection and Smart or Smart PRO controllers. For biological material and products requiring ultra-low temperatures.

ZLN-UT 200 VIP

ZLN-UT 200 VIP

Chamber volume
259 l
Temperature
-86…-50 °C
ZLN-UT 300 VIP

ZLN-UT 300 VIP

Chamber volume
345 l
Temperature
-86…-50 °C
ZLN-UT 500 VIP

ZLN-UT 500 VIP

Chamber volume
482 l
Temperature
-86…-50 °C

300 VIP and 500 VIP capacities are per manufacturer data; available versions and controllers are confirmed on enquiry.

0…+15 °C · Zone 2

ATEX laboratory refrigerators

CHL refrigerators for safe storage of samples that may create an explosive atmosphere, built to ATEX Directive 2014/34/EU, marking II 3/- G Ex h IIC T6 Gc/-. Zone classification is defined by the employer through risk assessment.

CHL 1 ATEX

CHL 1 ATEX

Chamber volume
74 l
Working capacity
54 l
Temperature
0…+15 °C
Specification
CHL 2 ATEX

CHL 2 ATEX

Chamber volume
157 l
Working capacity
129 l
Temperature
0…+15 °C
Specification
CHL 3 ATEX

CHL 3 ATEX

Chamber volume
208 l
Working capacity
171 l
Temperature
0…+15 °C
Specification
CHL 4 ATEX

CHL 4 ATEX

Chamber volume
259 l
Working capacity
212 l
Temperature
0…+15 °C
Specification
CHL 5 ATEX

CHL 5 ATEX

Chamber volume
312 l
Working capacity
255 l
Temperature
0…+15 °C
Specification
CHL 6 ATEX

CHL 6 ATEX

Chamber volume
414 l
Working capacity
339 l
Temperature
0…+15 °C
Specification

Emergency Power Supply

A dedicated battery backup system for refrigerators, thermostatic cabinets, cooled incubators and freezers during power outages. It helps maintain the cold chain of medicines and vaccines for outages of 4 to 30 hours. A classic computer UPS is not sufficient because the compressor draws a multiple of its rated current at start-up; POL-EKO units tolerate 300% overload for 20 s. Standard: inverter with charger, battery, braked casters, visual and audible signalling, type E (230 V) socket.

POL-EKO ZA II 4H

ZA II 4H

Nominal runtime
4 h
Compatible with
CHL 500–1450, ST 500–1450, ILW 240–750
Specification
POL-EKO ZA II 8H

ZA II 8H

Nominal runtime
8 h
Compatible with
CHL 500–1450, ST 500–1450, ILW 240–750
Specification
POL-EKO ZA II 12H

ZA II 12H

Nominal runtime
12 h
Compatible with
CHL 500–1450, ST 500–1450, ILW 240–750
Specification
Chamber combinations

Multi-chamber units

Double and triple-chamber CHL units and combinations with ST thermostatic cabinets and a ZLN 85 freezer in one housing. Multi-chamber units must be anchored to a wall.

CHL 1/1/1

CHL 1/1/1

Chamber volume
74 l
Working capacity
54 l
Temperature
0…+15 °C
Specification
CHL 2/2

CHL 2/2

Chamber volume
157 l
Working capacity
129 l
Temperature
0…+15 °C
Specification
CHL 2/3

CHL 2/3

Chamber volume
157 l
Working capacity
129 l
Temperature
0…+15 °C
Specification
CHL2/ST2

CHL2/ST2

Chamber volume
157 l
Working capacity
129 l
Temperature
0…+15 / +3…+70 °C
Specification
CHL3/ST2

CHL3/ST2

Chamber volume
208 l
Working capacity
171 l
Temperature
0…+15 / +3…+70 °C
Specification
CHL2/ST3

CHL2/ST3

Chamber volume
157 l
Working capacity
129 l
Temperature
0…+15 / +3…+70 °C
Specification
CHL2/ZLN85

CHL2/ZLN85

Chamber volume
157 l
Working capacity
129 l
Temperature
0…+15 / -25…0 °C
Specification
CHL3/ZLN85

CHL3/ZLN85

Chamber volume
208 l
Working capacity
163 l
Temperature
0…+15 / -25…0 °C
Specification
+20…−22 °C · 25–55 l

Dometic CFX5 transport refrigerators

Portable compressor coolers for moving samples, reagents and temperature-sensitive goods in the field and in a vehicle. They run on 12/24 V DC from a vehicle and 100–240 V AC, with VIP insulation and app-based temperature monitoring. They are intended for general use: if transport is part of a validated cold chain, temperature must be qualified and mapped separately under real conditions.

Cold chain validation

From refrigerator to evidence.

The unit is only the first step. For medicines, vaccines and temperature-sensitive samples it matters to prove that the cold chain stays within the specified limits, and that proof is built through mapping and qualification.

COLD CHAIN

Uninterrupted temperature control

Storage, transport and dispensing without leaving the specified range, supported by alarms, data logging and a response to deviations. A refrigerator with a microprocessor controller and alarms is the foundation of the chain.

MAPPING

Temperature distribution in the chamber

Measuring temperature at multiple points in an empty and a loaded chamber reveals the warmest and coldest spots, behaviour on door opening and power failure, and a suitable placement of sensors and shelves.

QUALIFICATION

IQ · OQ · PQ

Installation, operational and performance qualification documents that the unit performs to user requirements under real conditions. The results are part of cold chain process validation.

What the path to validation looks like.

Scope and execution depend on user requirements, internal procedures and applicable regulations. APICAL helps with planning, while measurements and documentation are carried out by qualified specialists.

POL-EKO offers Smart and Smart PRO controllers (LAN, USB, plus Wi-Fi on Smart PRO), independent temperature recorders with SMS notification, LabDesk software (including a CFR 21 Part 11 compliant version) and calibration in its own accredited measurement laboratory. Option availability is confirmed per model. The above are general guidelines; actual requirements are defined by applicable regulations (e.g. GDP/GMP) and your quality system.

  1. 1
    User requirements

    Define the goods, temperature range, capacity and criticality of storage (URS).

  2. 2
    Installation and IQ

    Place the unit and verify connections, sensors, alarms and documentation.

  3. 3
    Mapping and OQ/PQ

    Temperature mapping of the empty and loaded chamber, alarm tests, door opening and power failure tests.

  4. 4
    Monitoring and calibration

    Continuous recording and alarming, periodic calibration and requalification as planned.

Choosing cooling equipment?

Let's match the unit to your requirements.

Tell us what you store, the required temperature, capacity and validation requirements. The APICAL team can help you review suitable POL-EKO models.

Send an enquiry ↗