HomeProductsThermostatic equipment
Thermostatic equipment · POL-EKO

Controlled conditions, repeatable results.

Incubators, drying ovens and climatic chambers for storing and testing samples at precisely set temperature, humidity and light. APICAL helps match the choice to application, capacity and validation requirements.

Mapping and qualification as part of thermostatic equipment validation
POL-EKO incubator, drying oven and climatic chamber
POL-EKOIncubators, drying ovens and climatic chambers

A test result is only as reliable as the conditions in the chamber. Uniform temperature, documented validation and the right choice of unit go together.

APICAL × POL-EKOA presentation of device families, capacities and temperature ranges.
+3…+100 °C · 14–1479 l

Incubators

For incubating samples in a stable environment. From simple heated incubators to cooled, Peltier and CO₂ versions, with Smart and Smart PRO controllers that program time-temperature profiles.

POL-EKO CLN

Laboratory incubators CLN

Natural air convection without a fan, suited to delicate samples that quickly lose moisture.

Working capacity
14–237 l
Temperature
ambient +5…+100 °C
Models
6 models: 15–240
Specifications
POL-EKO CLW

Laboratory incubators CLW

Forced air circulation for excellent uniformity and quick temperature recovery after door opening.

Working capacity
14–998 l
Temperature
ambient +5…+100 °C
Models
9 models: 15–1000
Specifications
POL-EKO ILW

Cooled incubators ILW

Heating and cooling for a stable temperature regardless of room conditions, with double door and lock.

Working capacity
55–749 l
Temperature
0…+70 °C (−10 °C option)
Models
5 models: 53–750
Specifications
POL-EKO ST

Cooled incubators ST

For stability testing and storing medicines and samples at +18…+20 °C. Multi-chamber versions have independently controlled chambers.

Working capacity
54–1479 l
Temperature
+3…+40 °C (up to +70 °C option)
Models
ST 1–1450, 2- and 3-chamber
Specifications
POL-EKO ILP / STP

Peltier incubators ILP and STP

Peltier cell cooling without a compressor or refrigerant: vibration-free, quiet and compact. Cooling up to 20 °C below ambient.

Working capacity
55–749 l (ILP), 222 l (STP 4)
Temperature
0…+70 °C
Models
ILP 53–750, STP 4
Specifications
POL-EKO ILC

CO₂ incubators ILC

New in the POL-EKO range: precise control of temperature, humidity and CO₂ concentration for cell culture work.

Working capacity
180 / 260 l
Temperature
ambient +5…+50 °C
Models
ILC 180, ILC 260
Specifications
up to +300 °C · 14–998 l

Drying ovens

Ovens for drying and heat treatment with natural or forced air circulation and a Smart, Smart PRO or simple SIMPLE controller. Nitrogen blow versions, pass-through sterilizers and CALDERA warmers are also available.

POL-EKO SLN

Drying ovens SLN

Natural air convection with Smart or Smart PRO controllers. For drying and heat treating delicate materials.

Working capacity
14–237 l
Temperature
ambient +5…+300 °C
Models
7 models: 15–240
Specifications
POL-EKO SLW

Drying ovens SLW

Forced air circulation for faster and more uniform heating, from benchtop units to 1000-litre ovens.

Working capacity
14–998 l
Temperature
ambient +5…+300 °C
Models
10 models: 15–1000
Specifications
POL-EKO SIMPLE

SIMPLE drying ovens

A simple controller that programs temperature only. A practical choice when advanced programming is not needed.

Working capacity
55 / 107 l
Temperature
ambient +5…+250 °C
Models
SLN and SLW SIMPLE 53, 115
Specifications
POL-EKO SLWN

Drying ovens with nitrogen blow SLWN

For drying samples at 105–110 °C with nitrogen flow per ISO 589. Supplied with connections, valves and a rotameter; cylinder not included.

Working capacity
14–237 l
Temperature
ambient +5…+300 °C
Models
5 models: 15–240
Specifications
POL-EKO SRWP

Pass-through sterilizers SRWP

A link between clean and dirty zones on production lines: stainless steel chamber, forced air circulation and a Smart touch display.

Working capacity
107 / 245 l
Temperature
ambient +5…+250 °C
Models
SRWP 115, SRWP 240
Specifications
POL-EKO CALDERA

Blanket and fluid warmers CALDERA

Safe warming of fluids and blankets with alarms, over-temperature protection and parameter logging. Complies with PN-EN 60601-1-2.

Working capacity
61–268 l
Temperature
+35…+42 °C (+35…+70 °C TERM)
Models
5 models: 70–300
Specifications
0…+100 °C · 10…90 % RH

Climatic chambers

Specialised chambers for ageing, stability, thermal and climatic tests, with humidity control and, in phytotron and ICH versions, light control. Built from acid-resistant steel with double doors and lock.

POL-EKO KKS

Climatic chambers KKS

Steam humidifier, 7" Smart PRO controller (Wi-Fi, LAN, USB), automatic defrost and double doors. For stability testing per ICH Q1A.

Working capacity
107–1326 l
Temperature
0…+100 °C · 10…90 % RH
Models
8 models: 115–1450
Specifications
POL-EKO KKP

Climatic chambers KKP

Peltier technology: quiet, vibration-free and refrigerant-free, nearly maintenance-free. For ICH Q1A stability, plants and insects.

Working capacity
237 / 749 l
Temperature
0…+70 °C · 10…90 % RH
Models
KKP 240, KKP 750
Specifications
POL-EKO KKS FIT

Phytotron chambers KKS FIT

The only chambers that control temperature, humidity and lighting together, with LED day/night simulation (FIT P, S, D, DS).

Working capacity
107–1326 l
Temperature
0…+60 °C (+10…+50 °C with light)
Models
8 models: 115–1450
Specifications
POL-EKO KKS ICH

Photostability chambers KKS ICH

ICH Q1B photostability testing in just 29 hours for pharmaceuticals, cosmetics and packaging, with LabDesk CFR software (21 CFR Part 11).

Working capacity
215 / 702 l
Temperature
+10…+60 °C (ICH Q1A)
Models
KKS 240 ICH, KKS 750 ICH
Specifications
Validation

From the unit to the evidence.

For incubators, ovens and chambers, reliability is proven by mapping temperature (and humidity) in the chamber and by IQ, OQ and PQ qualification. The results form part of validation of the process the unit supports.

STABILITY

ICH Q1A and Q1B

KKS, KKP and KKS ICH climatic chambers are intended for stability testing of medicines at controlled temperature, humidity and light.

MAPPING

Condition distribution in the chamber

Multi-point measurement of the empty and loaded chamber reveals the hottest and coldest spots and behaviour on door opening and power failure.

QUALIFICATION

IQ · OQ · PQ

Installation, operational and performance qualification documents that the unit performs to user requirements. LabDesk software support is available, including a 21 CFR Part 11 compliant version.

What the path to validation looks like.

Scope and execution depend on user requirements, internal procedures and applicable regulations. APICAL helps with planning, while measurements and documentation are carried out by qualified specialists.

POL-EKO offers Smart and Smart PRO controllers (LAN, USB, plus Wi-Fi on Smart PRO), LabDesk software and calibration in its own accredited measurement laboratory. Option availability is confirmed per model; capacities are working volumes per manufacturer data. The above are general guidelines; actual requirements are defined by applicable regulations (e.g. GMP, ICH) and your quality system.

  1. 1
    User requirements

    Define the samples, temperature range, humidity, capacity and process criticality (URS).

  2. 2
    Installation and IQ

    Place the unit and verify connections, sensors, alarms and documentation.

  3. 3
    Mapping and OQ/PQ

    Mapping of the empty and loaded chamber, alarm tests, door opening and power failure tests.

  4. 4
    Monitoring and calibration

    Continuous recording and alarming, periodic calibration and requalification as planned.

Choosing thermostatic equipment?

Let's match the unit to your requirements.

Tell us which samples you process, the required temperature and humidity, capacity and validation requirements. The APICAL team can help you review suitable POL-EKO models.

Send an enquiry ↗